Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 70 III: (1 year) CAL VER /5520 DATE: 03-Dec-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.2 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 20 NUMBER OF LINES: 106 CONFIGURATION: Fluke 5520A ============================================================================= # # Source: # Fluke 70,73,75,77,21 and 23 Series III Service Manual # (PN 659141 November 1997) # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later # # Subprocedures: # None # # Required Files: # 7x_dsp.bmp # 52_70_cw.bmp # 52_70_2w.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5520A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R P F W 1.002 ASK+ K 1.003 HEAD EQUIPMENT SETUP 1.004 DISP [32] WARNING 1.004 DISP HIGH VOLTAGE is used or exposed during the performance 1.004 DISP of this calibration. DEATH ON CONTACT may result if 1.004 DISP personnel fail to observe safety precautions. 1.005 DISP Allow the UUT to stablize to room temperature 1.005 DISP 23degC +/-5degC (73degF +/-9degF). 1.005 DISP 1.005 DISP Check the fuses and battery, and replace them if 1.005 DISP necessary. 1.006 HEAD {DISPLAY TEST} 1.007 PICE 7x_dsp 2.001 HEAD {RESISTANCE TESTS} 2.002 DISP Select the Ohms Function. 2.003 PIC 52_70_cw 2.004 HEAD RESISTANCE TESTS: {320 Ohm Range} 2.005 5520 320 0.0Z 0.2U CW 3.001 5520 320 270.0Z 1.6U CW 4.001 HEAD RESISTANCE TESTS: {3200 Ohm Range} 4.002 5520 3200 2700Z 15U CW 5.001 HEAD RESISTANCE TESTS: {32 kOhm Range} 5.002 5520 32 27.00kZ 0.15U CW 6.001 HEAD RESISTANCE TESTS: {320 kOhm Range} 6.002 PIC 52_70_2W 6.003 5520 320 270.0kZ 1.5U 2W 7.001 HEAD RESISTANCE TESTS: {3.2 MOhm Range} 7.002 5520 3.2 2.700MZ 0.015U 2W 8.001 HEAD RESISTANCE TESTS: {32 MOhm Range} 8.002 5520 32 27.00MZ 0.55U 2W 9.001 HEAD {DIODE TEST} 9.002 M5520 RNGLK 33V 9.003 DISP Select the Diode Test Function. 9.004 5520 0.090V S 2W 9.005 EVAL Is the {beeper on}? 10.001 5520 0.110V S 2W 10.002 EVAL Is the {beeper off}? 11.001 M5520 * 11.002 HEAD {DC MILLIVOLT TEST} 11.003 DISP Select the 300mV DC Function. 11.004 HEAD DC MILLIVOLT TEST: {320mV Range} 11.005 5520 350 300.0mV 1.0U 2W 12.001 HEAD {DC VOLTAGE TESTS} 12.002 DISP Select the V DC Function. 12.003 HEAD DC VOLTAGE TESTS: {3.2V Range} 12.004 5520 3.2 2.700V 0.009U 2W 13.001 HEAD DC VOLTAGE TESTS: {32V Range} 13.002 5520 32 27.00V 0.09U 2W 14.001 HEAD DC VOLTAGE TESTS: {320V Range} 14.002 5520 320 270.0V 1.5U 2W 15.001 HEAD DC VOLTAGE TESTS: {600V Range} 15.002 5520 600 600V 3U 2W 16.001 HEAD {AC VOLTAGE TESTS} 16.002 DISP Select the V AC Function. 16.003 HEAD AC VOLTAGE TEST: {3.2V Range} 16.004 5520 3.2 2.700V 0.056U 100H SI 2W 17.001 5520 3.2 2.700V 0.056U 500H SI 2W 18.001 HEAD AC VOLTAGE TEST: {600V Range} 18.002 5520 600 600V 14U 100H SI 2W 19.001 5520 600 600V 14U 1kH SI 2W 20.001 END